52 mm x 32 mm LCD (Liquid Crystal Display), 5 digits.
|
|
|
dB (A & C frequency weighting), Time weighting(Fast, Slow), Hold, Memory ( max. & min. ), Peak hold, AC & RS232 output.
|
|
|
|
|
|
|
|
|
Frequency weighting meet IEC 61672 class 2, calibrating input signal on 94 dB ( 31.5 Hz to 8 KHz), then the accuracy of frequency weighting is specified as following:
31.5Hz-±3dB, 63Hz-±2.5dB, 125Hz-±2.0dB 250Hz-±1.9dB, 500Hz-±1.9dB, 1kHz-±1.4dB 2kHz-±2.6dB, 4kHz-±3.6dB, 8kHz-±5.6dB
|
Frequency Weighting Network
|
Characteristics of A & C.
|
A weighting – The characteristic is simulated as “Human Ear Listing” response. Typical, if making the environmental sound level measurement, always select to A weighting.
|
C weighting – The characteristic is near the “FLAT” response. Typical, it is suitable for checking the noise of machinery (Q.C. check) & knowing the sound pressure level of the tested equipment.
|
|
|
|
|
|
B & K (Bruel & kjaer), multi-fuction acoustic calibrator, model : 4226.
|
|
|
Electric condenser microphone.
|
|
|
|
|
|
Auto range : 30 to 130 dB.
|
Manual range : 3 range, 30 to 80 dB, 50 to 100 dB, 80 to 130 dB, 50 dB on each step, with over & under range indicating.
|
Time Weighting (Fast & Slow)
|
Fast – t= 200 ms, Slow – t = 500 ms,
* “Fast” range is simulated the human ear response time weighting.
“Slow” range is easy to get the average values of vibration sound level.
* The “Fast” & “Slow” time weighting range are designed to IEC 61672 class 2 requirement.
|
|
|
* AC output – AC 0.5 Vrms corresponding to each range step. Output impedance – 600 ohm.
* RS232 output.
|
|
|
Terminal 1 : RS232 computer interface terminal. Terminal 2 : AC output terminal.
* Terminal socket size : 3.5 mm dia. phone socket.
|
|
|
Build in external calibration VR, easy to calibrate on 94 dB level by screw driver.
|
|
|
0 °C to 50 °C ( 32°F to 122°F ).
|
|
|
|
|
|
006P DC 9V battery ( Alkaline or heavy duty type ).
|
|
|
|
|
|
|
205 x 68 x 29 mm ( 8.07×2.7×1.1 inch ).
|
|
|
170.5 x 24.5 x 19 mm ( 6.7×1.0x0.7 inch ).
|
|
|
405 g/0.89 LB ( including probe ).
|
|
|
Instruction manual …………………………….. 1 PC.
Separate sound probe…………………………. 1 PC.
Probe holder……………………………………… 1 PC.
|
|
|
Sound Calibrator, model : SC-934 ( 94/104/114 dB ).
SC-942 ( 94dB, 114 dB ).
Carrying case : CA-06
RS232 cable, Model : UPCB-02
USB cable, Model : USB-01
Application software, Model : SW-U801-WIN
|
There are no reviews yet.