One Channel

Signal Conditioner with Direct Current Excitation and a gain of 10000 times maximum.

KYOWA CDV, CDA-700A

More details

Call 022 7106162

Add Subtract

CDV-700A
CDA-700A
8 Units Accommodated in Portable Housing Case YB-508A

 

Features

  • Frequency response range is as wide as DC to 500 kHz, suitable for high-speed phenomena such as impulse wave.
  • Strain amplifier mode provides a range of 100μm/m.
  • Noise referred to input is low despite the wide bandwidth and high gain capabilities.
  • Balance adjustment is possible up to ±10000μm/m (equivalent to a resistance deviation of 2%).
  • RANGE, L.P.F., VERN and BAL settings are saved in nonvolatile memory.
  • User registration function improves the measurement accuracy. such as ;
    1) With CDV-700A, the cable resistance compensation function ensures the same measurement accuracy as ensured by remote sensing.
    2) With CDA-700A, other than 120 Ω or 350Ω .
  • Output of 4 to 20 mA is option.
  • Digital indicator and bar graph are provided.
  • Powerful low-pass filter provides the attenuation characteristic of -24 dB / oct.
  • Key lock function prevents maloperation of balance, range, vernier and low-pass filter switches.

PDF download

 

Model

ModelCDV-700ACDA-700A
Bridge Excitation ModeConstant voltageConstant current
Applicable BridgeResistance60 to 1000Ω

120 Ω, 350 Ω

User registration functionallows for selection ina range of 60.0 to 9999Ω.

Bridge Voltage(selectable by rearpanel dip switch)

10VDC

4VDC

2VDC

1VDC
(30 mA orless foreach voltage)

350Ω 25m

A350Ω 5mA

120Ω 25mA

120Ω 5mA

Sensitivity(to 100 m/m input)BVOutputBRBCOutput
10VDC
4VDC
2VDC
1VDC
5V
2V
1V
0.5V
350Ω
350Ω
120Ω
120Ω
25mA
5mA
25mA
5mA
4.375V
0.875V
1.5V
0.3V
Setting Rangeby User RegistrationFunction

Cable resistancein one way

0.0 to 200.0Ω

Bridge resistance

60.0 to 999.9Ω

or 60 to 9999Ω

Cable ExtensionRemote sensing enables up to 2 kmby 6-conductor (0.5 mm2) shielded cableUp to 2 kmby 4-conductor (0.5 mm2) shielded cable

 

Common Specifications

Number of Measuring Channels1 (multiple units can be combined formultichannel configuration)
Gage Factor2.00 fixed
Input Impedance10MΩ +10MΩ or more
Balance Adjustment Range (BAL)Resistance ±2%
(Initial unbalance of transducer 10000μm/m)
Balance Adjustment MethodResistance : True electron auto balane
Accuracy : ±2 μm/m (with range set to 100μm/m)Saved in nonvolatile memory
Output (Dual output)
OUTPUT A±10 V or more (load 5 KΩ or more)
Zero adjustment range ±0.1 V or more
OUTPUT B±10 V or more (load 5 KΩ or more)
Sensitivity adjustment range 1 to 1/100
Zero adjustment range ±30 mV or more to OUTPUT A
NonlinearityWithin ±0.01% FS
Output ImpedanceApprox. 2Ω
Sensitivity Switch (RANGE)
Strain amplificaton7 steps of 100, 200, 500, 1k, 2k, 5kμm/m]and OFF
DC amplification 7 steps of 10000, 5000, 2000,1000, 500, 200 times and OFF
AccuracyWithin ±0.1%
Saved in nonvolatile memory
Fine Sensitivity Control (VERN)1 to 1/2.5 or more
Saved in nonvolatile memory
Calibration (CAL)
Reference equivalent strain±(1 to 9999μm/m )
DC voltage±(10 to 99990 μV)
Setting4-digit switch
Accuracy±(0.2% + 0.5μm/m ),(0.1% + 5.0μ V RTI)
Frequency Response RangeDC to 500 kHz, deviation +1/-3 dB
Low-Pass Filter (L.P.F)4-pole fourth-order Butterworth
5 steps of 10, 100, 1k, 10kHz and F (flat)
Saved in nonvolatile memory
Amplitude ratio at cutoff point: -3 dB±1 dB
Attenuation: -24±1 dB/oct.

Noise

Low-pass FilterNoise (RTI)
10Hz3 μV p-p or less
100Hz4 μV p-p or less
1kHz6 μV p-p or less
10kHz12 μV p-p or less
Pass60 μV p-p or less

1 μV p-p RTI = 1 μm/m, with bridge voltage 2 V

Maximum Allowable Input Voltage±15V
Allowable Common Mode Input Voltage±10V
Common Mode Rejection Ratio100 dB or more

Zero Stability

 

With bridge voltage 2 V ±1μm/m/℃,± 8 μm/m/24 h
±0.05% FS±10% power fluctuation

Sensitivity Stability

 

±0.01%/℃, ±0.05%/24 h
±0.05%/±10% power fluctuation
Withstand Voltage1000 VAC for one minite between AC power
supply and chassis
Over-input IndicationPanel meter flickers.
Output Voltage Indication4-digit LED display for OUTPUT A, updated
approximately 15 times per second
10-segment LED bar graph
Key Lock FunctionRANGE, BAL, VERN, CAL, L.P.F.
switches can be locked.
Remote FunctionAllows remote execution of zero balance,calibration and key lock
Vibration ResistanceConforms to MIL-STD-810℃. 10 to 500 Hz
19.61 μm/s 2 (2 G) 5 to 55 Hz 29.42 μm/s 2 (3 G)
12 cycles each in X, Y and Z directions,
15 minutes/cycle
Operating Temperature & Humidity Range– 10 to 50℃, 85% RH or less (noncondensing)
Storage Temperature– 30 to 70℃
Power SupplyAC line, approx. 12 VA (100 VAC)
or 10.5 to 15 VDC, approx. 0.8 A (12 VDC)
Dimensions & Weight49(W) x 128.5(D) x 262.5(H) mm (excluding protrusions), approx. 1.4 kg
Panel cut dimensions50 x 113 mm

Standard Accessories

AC power cable P-25 (with conversion adapter CM-39)

Output cable U-58

Spare fuse (Midget type 3A)

Miniature screwdriver, Instruction Manual

Optional Accessories

Housing cases YB-A ( here ), Amplifier stand FA-1B( here ), Input cable U-37 ( here ), Extension cables N-81 to N85 ( here )