CAM-60B, UCAM-65B
Data Loggers
Data Loggers
KYOWA UCAM-60B, UCAM -65B
Resolution as high as 0.1 μm/m
Call 022 7106162
Data Loggers
Data Loggers
KYOWA UCAM-60B, UCAM -65B
Resolution as high as 0.1 μm/m
Call 022 7106162
Can measure up to 20000 μm/m with a resolution as high as 0.1 μm/m. (※When using for Full bridge system)
The data logger UCAM-60B is an all-in-one measuring instrument developed in full pursuit of easier field measurement. Easy to operate keys, a bright readable display providing understandable presentation and a printer for immediate confirmation of measurement results. All these and more are incorporated in this compact unit to satisfy every need in field measurement. The UCAM-65B is a compact online data logger fully controlled through the PC. Both models can connect to, and simultaneously input signals from, strain gages, strain gage transducers, civil engineering transducers with temperature measuring function, potentiometer sensors, thermocouples and DC voltage-output instruments. They are also compatible with TEDS-installed sensors having information conforming to IEEE template No. 33. While measurement in a maximum 30 channels is possible with the mainframe only, external scanners enable measurement in a maximum 1000 channels. Measured results are stored in internal memory. And for easy data transfer to PC, measured results can also be saved in a flash ATA card or CF card inserted into the PC card slot. Furthermore, Ethernet LAN and RS-232C interfaces are provided standard for connection to the PC, and the control software UCS-60B enables the PC to not only control the UCAM-60B/65B but also perform data processing for rosette analysis, etc. in the field by directly collecting data.
【UCAM-60B】
【UCAM-65B】
【Common to UCAM-60B and UCAM-65B】
(Note 1) For TEDS, refer to P.9-15.
(Note 2) When connecting via Ethernet LAN use a straight cable and LAN hub
Data Logger(Main Unit)
Model | Power Supply | Control Software UCS-60B |
---|---|---|
UCAM-60B-AC | AC only | Optional |
UCAM-60B-DC | DC only | Optional |
UCAM-65B-AC | AC only | Standard |
UCAM-65B-AC-0 | Optional | |
UCAM-65B-DC | DC only | Standard |
UCAM-65B-DC-0 | Optional |
Dedicated Scanners | USS-61B for general purpose |
USS-62B for general purpose with NDIS connectors*1 | |
USS-63B for civil engineering with lightning arrester | |
The main unit can accommodate up to 3 dedicated scanners. | |
External Scanners | The main unit can connect to the following scanners via the optional scanner interface. |
USB-70 series via scanner interface USI-67A | |
Scanner Interfaces | USI-67A for USB-70 series |
External Input/Output Unit | UIO-60A |
Control Software | UCS-60B |
*1. TEDS-compatible function is made effective by connecting
TEDS-installed sensor through NDIS connector.
UCAM-60B、65B | |
Applicable Sensors | Strain gages, strain gage transducers, civil engineering transducers with temperature measuring function, DC voltage-output or DC current-output instruments, potentiometer sensors,thermal sensors (thermocouples and platinum resistance thermometer bulbs) |
Scanners and Applicable Sensors | Dedicated Scanner | Dedicated Scanner | |||
---|---|---|---|---|---|
General purpose | Civilengineering | ||||
USB-70B-10/20 | USB-70B-30 | ||||
strain gages and strain gage transducers | Quarter bridge method | 120Ω | ● | ● | ● |
240Ω | ● | ● | ● | ||
350Ω | ● | ● | ● | ||
Quarter bridge method (true-dummy method)
| 120Ω | ● | ● | ● | |
240Ω | ● | ● | ● | ||
Half bridge method 60 to 1000
| Active-dummy method | ● | ● | ● | |
Active-active method | ● | ● | ● | ||
Common dummy method | ● | ● | |||
Full bridge method 60 to 1000 *3
| Opposite side active method | ● | ● | ● | |
Full-bridge method | ● | ● | ● | ||
Civil engineering transducers | Full bridge method 120 | Constant-current excitation | ● | ||
Full bridge method 350
| Constant-current excitation | ● | ● | ● | |
With temp. measuring function | ● | ● | |||
Voltage | DC voltage- output instruments | ● | ● | ● | |
Current | DC current- output instruments | ● | ● | ● | |
Temperature | Thermocouples | K(CA) | ● | ● | ● |
T(CC) | ● | ● | ● | ||
E(CRC) | ● | ● | ● | ||
J(IC) | ● | ● | ● | ||
R | ● | ● | ● | ||
Platinum resistance thermometer bulbs
| Pt100 (new JIS) | ● | ● | ||
JPt100 (old JIS) | ● | ● | |||
Potentiometer sensors | ● | ● | ● | ||
Built-in lightning arrester | ●(*1) | ● | |||
Scanner interface *3 | Not required | -67A |
*1. With USS-63B mounted.
*2. Use either of the scanner interfaces for
*3. 120 to 1000Ω in high-resolution mode.
Number of Measuring Channels | Max. 30 with dedicated scanners |
Max. 1000 with external scanners connected | |
Scanning Speed | 50 ms/channel (standard mode) |
280 ms/channel (high-resolution mode), individually selectable for | |
desired channels 20 ms/channel (high-speed mode), | |
collectively selectable for all channels of dedicated scanners |
Line Frequency | 50 Hz Zone | 60 Hz Zone |
---|---|---|
Scanner | ||
Dedicated scanner (standard mode) | 50 ms/channel | |
Dedicated scanner (high-resolution mode) | 280 ms/channel | |
Dedicated scanner (high-speed mode) | 20 ms/channel | |
USB-70 series (standard mode only) | 60 ms/channel | 58.4 ms/channel |
Note : Scanning speeds stated above are standard maximum speeds in respective modes. Besides these, the following speeds can be set for each individual channel: 0.28 s, 0.5 s, 1 s, 2 s, 5 s and 10 s
Scanning Speed |
Standard Mode (50 ms/channel)
|
High-Resolution Mode (280 ms/channel)
|
High-Speed Mode (20 ms/channel)
|
---|---|---|---|
Applicable Sensor | |||
Strain (gage & transducer) | ● | ● | ● |
Voltage/current-output sensor | ● | × | ● |
Civil engineering transducer | ● | × | × |
Temperature sensor (TC, Pt) | ● | × | × |
Potentiometer sensor | ● | × | ● |
Notes : 1. High-resolution mode and high-speed mode are selectable for dedicated anners only.
2. High-resolution or high-speed mode is available only with
4-gage method.
Operating Modes | Real-time, |
Monitor, and Automatic | |
Measurement Functions | Initial: Initial values are measured and stored in internal memory (except for temperatures measured by civil engineering transducers with temperature measuring function). |
Original:Raw values are measured without subtraction of initial values. | |
Measure : Initial values are subtracted from original values (except for temperatures measured by civil engineering transducers with temperature measuring function). | |
Easy Measure : Auto zero balancing function is activated. | |
Note: The selected function is applied to all channels. Coefficient | |
Calculation Function | Multiplication by calibration coefficients,calibration by TEDS, conversion of measured values to physical quantities, |
scaling and correction. | |
Engineering Units | 59 units |
Automatic Measurement Functions | |
Trigger Measurement | A relative value (certain changing quantity) or an absolute value triggers measurement. In addition to the usual trigger function, a variable trigger function is provided with which the trigger value changes at each step during measurement. With this special function, a trigger value and the number of measurement times (repeat times) under the trigger condition can be registered for each step to perform a series of automatic measurements in the order of steps. The maximum number of steps available for setting is 15 and the number of repeat times may be a value selected from a range of 1 to 9999 or infinite. |
Trigger channel : 1 desired channel | |
Trigger value : A desired real number of 6 effective figures or less Reference | |
Reference value : Amount of level shift to determine the first trigger value | |
(selected from the same range as for the trigger values) | |
Number of repeat times : 1 to 9999 (0 for infinite times) | |
Number of measurement steps : Maximum 15 |
INTERVAL MEASUREMENT | Measurement is automatically |
performed at preset time intervals. | |
Starting time setting : Year/month/day and hour:minute:second | |
Interval setting : Days and hours:minutes:seconds (in a range | |
of 00 and 00:00:01 to 99 and 12:59:59) | |
Number of repeat times : 1 to 9999 (0 for infinite times) | |
Maximum number of steps : 15 |
TRIGGER INTERVAL MEASUREMENT | Combination of trigger measurement and interval measurement. |
Trigger value: Absolute value Number of measurement times: Max. 9999 | |
Interval time: Available in a range of 1 second to 99 days 23 hours:59 minutes:59 seconds | |
Storage Function | Internal memory,approx. 7 MB |
Flash ATA card (optional); the capacity depends on the card. | |
Strain Measurement(Standard Mode) | |
Bridge excitation | Constant voltage excitation : Approx. 2 or 5 VDC |
Constant current excitation : | |
Approx. 5.7 mA (bridge resistance 350Ω) | |
(up to 5 km with a 4-conductor (0.5 mm 2 ) shielded cable) | |
Approx. 16.7 mA (bridge resistance 120Ω) | |
(up to 2 km with a 4-conductor (0.5 mm 2 ) shielded cable) | |
Scanning speed | 50 ms/channel |
Gage factor | 2.00 fixed (coefficient calculation function enables 2.00 fixed (coefficient calculation function enables |
Initial value memory range | Same as measuring range |
Measuring range, Resolution and Accuracy
Measuring range, Resolution and Accuracy | Resolution | Accuracy |
---|---|---|
0 to ±50000μm/m | 1μm/m | ±(0.05% of reading + 1) μm/m |
±50000 to 500000μm/m | 10 μm/m | ±(0.05% of reading + 10)μm/m |
Strain Measurement (High-Resolution Mode)
Constant voltage excitation | Approx. 5 VDC |
Constant current excitation
| Approx. 16.7 mA (bridge resistance 350Ω) |
(up to 2 km with a 4-conductor (0.5 mm 2 ) shielded cable) | |
Scanning speed | 280 ms/channel |
Gage factor | 2.00 fixed (coefficient calculation function enables correction with 2.00/Ks) |
Initial value memory range | Same as measuring range |
Same as measuring range
Measuring Range | Resolution | Accuracy |
---|---|---|
0 to ±20000μm/m | 0.1μm/m | ±(0.05% of reading + 0.3) μm/m |
±20000 to 200000μm/m | 1μm/m | ±(0.05% of reading + 3) μm/m |
Notes : 1. Available only with 4-gage method (bridge resistance 120 to 1000Ω)
2. Bridge resistance should be 350Ω for bridge excitation with constant current.
3. Measuring range is 0 to 15000 μm/m for bridge excitation with constant current.
4. Available only with dedicated scanners.
Strain Measurement(High-Speed Mode) | |
Bridge excitation
| Constant voltage excitation: Approx. 2 VDC |
Constant current excitation: | |
Approx. 5.7 mA (bridge resistance 350Ω ) | |
(up to 5 km with a 4-conductor (0.5 mm 2 ) shielded cable) | |
Approx. 16.7 mA (bridge resistance 120Ω ) | |
(up to 2 km with a 4-conductor (0.5 mm 2 ) shielded cable) | |
Scanning speed | 20 ms/channel |
Gage factor | 2.00 fixed (coefficient calculation function enables correction with 2.00/Ks) |
Initial value memory range | Same as measuring range |
Measuring range, Resolution and Accuracy
Measuring Range | Resolution | Accuracy |
---|---|---|
0 to ±50000 μm/m | 1 μm/m | ±(0.08% of reading + 3) μm/m |
±50000 to 500000μm/m | 10μm/m | ±(0.08% of reading + 30) μm/m |
Notes : 1. Available only with 4-gage method
2. Available only with dedicated scanners.
Voltage Measurement (Stondard Mode) | |
Scanning speed | 50 ms/channel |
Initial value memory range | Same as measuring range |
Measuring range, Resolution and Accuracy
Range Mode | Measuring Range | Resolution | Accuracy | Input Resistance |
---|---|---|---|---|
V/500mV | 0 to ±50.000 mV | 1μV | ±(0.05% of reading + 3) |
10 M Ω or more
|
±50.00 to 500.00 mV | 10μV | |||
V/50V | 0 to ±5.0000 V | 100μV | ±(0.05% of reading + 2) |
1 M Ω or more
|
±5.000 to 50.000 mV | 1mV |
Voltage Measurement (High-Speed Mode) | |
Scanning speed | 20 ms/channel |
Initial value memory range | Same as measuring range |
Measuring range, Resolution and Accuracy
Range Mode | Measuring Range | Resolution | Accuracy | Input Resistance |
---|---|---|---|---|
V/500mV | ±(0.08% of reading + 6) | 1μV | ±(0.08% of reading + 6) |
10 M Ω or more
|
±50.00 to 500.00 mV | 10μV | |||
V/50V | 0 to ±5.0000 V | 100μV |
1 M Ω or more
| |
±5.000 to 50.000 mV | 1mV |
Current Measurement (Standard Mode) | |
Scanning speed | 50 ms/channel |
Initial value memory range | Same as measuring range |
Measuring range, Resolution and Accuracy
Channel Mode | Measuring Range | Resolution | Accuracy |
---|---|---|---|
I/50mA | 0 to ±50.00mA | 10μA | (±0.05% of reading + 0.01) mA |
Notes : 1. External shunt resistor (high-accuracy 250Ω ) is required.
2. Stated accuracy does not include the external shunt resistor.
Current Measurement (High-Speed Mode) | |
Scanning speed | 20 ms/channel |
Initial value memory range | Same as measuring range |
Measuring range, Resolution and Accuracy
Channel Mode
| Measuring Range | Resolution | Accuracy |
---|---|---|---|
I/50mA | 0 to ±50.00 mA | 10μA | ±(0.08% of reading + 0.01) mA |
Notes : 1. Available only with dedicated scanners.
2. External shunt resistor (high-accuracy 250Ω) is required.
3. Stated accuracy does not include the external shunt resistor accuracy.
Temperature Measurement with Thermocouples (Standard Mode) | |
Scanning speed | 50 ms/channel |
Measuring range,Resolution and Accuracy
Type | Measuring Range | Resolution | Accuracy | Internal Reference Junction Compensator Accuracy |
---|---|---|---|---|
K | -200.0 to 1230.0 ℃ | 0.1℃ | ±0.7℃ |
±0.5 ℃ (with input terminal temperature balanced in an ambient )(temperature range of 0 to 50 ℃)
|
T | -200.0 to 400.0 ℃ | ±0.7℃ | ||
E | -200.0 to 660.0 ℃ | ±0.5℃ | ||
J | -200.0 to 870.0 ℃ | ±0.6℃ | ||
R | -0 to 1760.0 ℃ | ±2.2℃ |
Notes : 1. Accuracies do not include the internal reference junction
compensator accuracy.
2. The reference junction compensator is switchable between internal
and external.
3. Thermocouple resistance should be 1 kΩor less.
Temperature Measurement with Civil Engineering Transducers with Temperature Measuring Function (Standard Mode) | |
Scanning speed | 50 ms/channel |
Measuring range, Resolution and Accuracy
Measuring Range | Resolution | Accuracy |
---|---|---|
-50.0 to 200.0℃ | 0.1℃ | ±0.5℃ |
Notes : 1. Target physical quantity and temperature can be measured
in a single channel.
2. Strain measuring ranges are the same as in strain measurement
in standard mode.
Temperature Measurement with Platinum Resistance Thermometer Bulb (Standard Mode) | |
Scanning speed | 50 ms/channel |
Measuring range, Resolution and Accuracy
Type | Measuring Range | Resolution | Accuracy |
---|---|---|---|
Pt100 | -200.0 to 660.0℃ | 0.1℃ | ±0.3℃ |
JPt100 | -200.0 to 510.0℃ |
Note : Connection is 3-wire system
Measurement with Potentiometer Sensor | |
Scanning speed | 50 ms/channel (standard mode) 20 ms/channel (high-speed mode) |
Initial value memory range | Same as measuring range |
Sensor power supply | Approx. 2 VDC |
Potentiometer resistance | 1 to 10 kΩ |
Measuring range, Resolution and Accuracy
Channel Mode | Measuring Range | Resolution | Accuracy |
---|---|---|---|
POT. | 0 to ±50.00% | 0.01% | ±0.1%FS |
Note : Connection is 3-wire system
Clock (UCAM-60B) | Real-time clock is built in (battery backup 5 years). |
Display (UCAM-60B) | Fluorescent display tube, |
128 x 64 dots Printer (UCAM-60B) | |
Printing system | Thermal |
Paper width : 58 mm (24 characters/line), UCAM-60A-RP | |
Printing speed : 60 mm/sec | |
PC Card Slot | Conforms with PCMCIA Ver. 4.2. |
Accepts a commercially available flash ATA card or | |
CF card (ATA card adapter required). | |
Interfaces | RS-232C and LAN (10BASE-T/100BASE-TX) |
File Conversion | CSV conversion |
Self-diagnosis Function | Checks display, printer, bridge excitation, leadwire-off, input/output resistance, insulation resistance, mode, etc. Checking of input/outut resistance and mode is availabe only for dedicated scanners. |
TEDS Compatibility | Interface : IEEE 1451.4 Mixed Mode Transducer Interface Class 2 |
Applicable sensor : Should have information written in accord-ance with IEEE template No. 33; cable length should be 30 m or less. | |
Operating Temperature & Humidity Range: | 0 to 50℃, 20 to 85% RH (noncondensing) |
Power Supply | AC85 to 264V, 50/60Hz (AC-operated version) |
10 to 16 VDC (DC-operated version) | |
Note : DC-operated version has power control function. | |
Current Consumption | 0.5 A or less: 100 VAC (with 3 dedicated scanners mounted) |
4 A or less: 12 VDC (with 3 dedicated scanners mounted) | |
Demensions | UCAM-60B: 360(W) x 88(H) x 400(D) mm (excluding protrusions) |
UCAM-65B: 327(W) x 88(H) x 365(D) mm (excluding protrusions) | |
Weight | UCAM-60B: Approx. 8 kg,UCAM-65B: Approx. 4.6 kg |
AC power cable P-18 with conversion adapter CM-39 (AC-operated version) DC power cable P-57 (DC-operated version)
Recording paper UCAM-60A-RP (1 roll for UCAM-60B only)
Screwdriver, Spare fuse, Instruction Manual
CD-R (Control Software UCS-60B for UCAM-65B only)
Recording Paper UCAM-60A-RP (10 rolls/pack)
USS-61B/62B/63B | |
Models | USS-61B (TEDS compatible) |
USS-62B (with NDIS connectors, TEDS compatible) | |
USS-63B (for civil engineering measurement,TEDS compatible, with lightning arrester)
| |
Number of Measuring Channels | 10/unit |
Switching Terminals | Semiconductor relays |
Input Terminals | Can connect to leadwires through either soldering or screwing. |
NDIS connectors (USS-62B) | |
One-touch terminal block JT-1A (optional) | |
Lightning Arrester | Provided (USS-63B only) |
Operating Temperature & Humidity Range | 0 to 50℃, 20 to 85% RH (noncondensing) |
Dimensions | 320(W) x 28(H) x 80(D) mm (excluding protrusions) |
Weight | USS-61B : Approx. 800 g (including terminal cover) |
USS-62B : Approx. 1 kg (including terminal cover) | |
USS-62B : Approx. 1 kg (including terminal cover) |
Terminal cover, Channel label and for USS-62B, NDIS connector caps (pre-attached to connectors)
USI-67A | |
Connectable Scanners | USB-70 series |
Number of Connectable Scanners | Max. 20 |
Operating Temperature & Humidity Range | 0 to 50℃, 20 to 85% RH (noncondensing) |
Dimensions | 99(W) x 50(H) x 163(D) mm (excluding protrusions), |
Weight | Approx. 160 g |
USI-67A
UIO-60A | ||
Output:
| ALARM signal | 4 channels (high/low limit checking) |
BUSY signal | 1 channel | |
Input:
| START signal | 1 channel |
STOP signal | 1 channel | |
RESET signal | 1 channel | |
RAINFALL signal | 1 channel | |
Operating Temperature & Humidity Range | 0 to 50℃, 20 to 85% RH (noncondensing) | |
Dimensions | 90(W) x 50(H) x 180(D) mm(excluding protrusions), | |
Weight | Approx. 140g |
USB-70B | ||
Models | USB-70B-10 (for general strain measurement) | |
USB-70B-20 (for general strain meas., with NDIS connectors) | ||
USB-70B-30 (for civil engineering, with lightning arrester) | ||
Number of Measuring Channels | 50/unit | |
Measuring Channel Mode | Selected for each channel from the mainframe | |
Connectable Sensors | USB-70B-10 | Strain gages, strain gage transducers,potentiometer Sensors,DC voltage-output instruments, thermocouples |
USB-70B-20 | Strain gages, strain gage transducers,potentiometer Sensors,DC voltage-output instruments, thermocouples(trans-ducers with NDIS connector can be connected) | |
USB-70B-30 | Strain gages, strain gage transducers,potentiometer Sensors,DC voltage-output instruments,thermal Sensors (thermo-couples, platinum resistancethermometer bulbs, civil engineering transducers withtemperature measuring function);lightning arrester built in | |
Power Supply | Supplied from data logger If the cable is extended | |
or if 4 or more scanners are connected, an optional | ||
UPS-70B should be mounted into scanners. | ||
UPS-70B operates on 90 to 132/180 to 264 VAC | ||
(no switchover required), 50/60 Hz | ||
Operating Temperature & Humidity Range | 0 to 50℃, 20 to 85% RH (noncondensing) | |
Dimensions | 302(W) x 107(H) x 500(D) mm (excluding protrusions) | |
Weight | Approx. 7.7 kg (USB-70B-30) | |
Approx. 8.5 kg (USB-70B-20) |