NTB-100A
NTB-201A
A revolutionary concept of measurement has emerged
Signals from a sensor are usually digitized through an amplifier and A/D converter via a bridge box, and the digital data is processed on a computer or a dedicated system. NTB-100A Series network terminal boxes revolutionize this process signals from a sensor are immediately digitized and fed to a computer for data processing.These NTBs not only streamline signal processing, but also contributes to eliminating extra cable thereby reducing cost of labor and installation.By placing an NTB near a sensor, only a single communication cable is required to build a wide area network (a total distance of 1 km), which is also useful because the digital transmission is hardly affected by noise. A single unit can measure 4 channels, and up to 8 units can be connected, for measurement of a maximum of 32 channels is possible.The NTB, which directly connects various sensors including strain gages, facilitates digital measurement in the field such as construction or building site, or for indoor experiments and researches.Handy logger SME-100A allows a wider range ofmeasurement with its.
Features
- The wide area, decentralized arrangement will be useful for the infrastructure of building and civil engineering.
- Network output is compliant to CAN, requiring a single wire to build the network.
- Measurement can be started immediately when the instrument is connected to a computer.
- Various ways of interlocking and connection are provided, broadening system applications.
- Compact, lightweight and affordable, allowing a small-sized system to be built on site easily.
- Digitizing data adjacent to the sensor allows noise resistant digital data to be transmitted.
PDF download
Specifications
Model | Bridge excitation | Sensor input terminal | 1-gage resistance |
---|
NTB-100A-120 | Constant-voltage | One-touch lock terminal | 120Ω |
NTB-101A-120 | Constant-voltage | Screw soldering terminal | 120Ω |
NTB-100A-350 | Constant-voltage | One-touch lock terminal | 350Ω |
NTB-101A-350 | Constant-voltage | Screw soldering terminal | 350Ω |
NTB-110A-350 | Constant-voltage | One-touch lock terminal | For 4-gage only |
NTB-111A-350 | Constant-voltage | Screw soldering terminal | For 4-gage only |
※Control Software NTB-10A Standard accessory
Bridge excitation | Applicable sensor | NTB model |
---|
General-purpose strain measurement | Civil engineering measurement |
---|
NTB-100A-120 NTB-101A-120 | NTB-100A-350 NTB-101A-350 | NTB-110A-350 NTB-111A-350 |
---|
NTB models andapplicable sensors | Strain gage Strain gage transducer | Quarter gage method | 120Ω | ○ | × | × |
350Ω | × | ○ | × |
Half bridge method 120 to 1000Ω | Active-active method | ○ | ○ | × |
Full bridge method120 to 1000Ω | Full bridge | ○ | ○ | × |
Constant current | Civi lengineering transducer | 4-gage method 350Ω | Full bridge method Civil engineering transducer with temperature | × | × | ○ |
× | × | ○ |
channels | 4 |
Scanning speed | Approx. 0.5 sec/channel for 0 to ± 30000 μm/m Approx. 1 sec/channel for ± 30000 μm/m or more With civil engineering transducer with temperature measuring function |
Bridge excitation | Approx. 2VDC for constant-voltage bridge excitation Approx. 5.6mA for constant-current bridge excitation (at bridge resistance 350Ω) |
Measuring range | Strain measurement 0 to ± 300000 μm/m (constant-voltage bridge excitation) 0 to ± 30000 μm/m (constant-current bridge excitation) Temperature measurement with civil engineering transducer with temperature measuring function -30.0 to 70.0℃ |
Resolution | Strain measurement 0 to ± 30000 μm/m : 1 μm/m ± 30000 to ± 300000 μm/m : 10 μm/m Temperature measurement with civil engineering transducer with temperature measuring function 0.1℃ |
Accuracy | Strain measurement 0 to ± 30000 μm/m : ± (0.05% rdg. + 2) μm/m ± 30000 to ± 300000 μm/m : ± (0.1% rdg. + 20) μm/m Temperature measurement with civil engineering transducer with temperature measuring function ± 0.5℃ |
TEDS | Read from TEDS sensors Operator input of channel names (Kyowa ID only) |
Power save mode | Provided ON/OFF using “OPT.3” DIP switch. |
interface | Dedicated interface conforming to CAN, cable extension up to 1km |
Operating temperature range | -10 to 50℃ |
Operating humidity range | 20 to 85%RH (no condensation) |
Power supply | 11 to 16VDC |
Current consumption (during operation at 12VDC) | Constant-voltage bridge excitation Operation: 100mA or less Standby : 60mA or less Standby (in power save mode) : 40mA or less Constant-current bridge excitation Operation : 70mA or less Standby : 60mA or less Standby (in power save mode) : 40mA or less |
Dimensions | One-touch lock type: 150 (W) × 28 (H) × 55 (D) mm (excluding protrusions) |
Screw soldering type | 150 (W) × 28 (H) × 110 (D) mm (excluding protrusions) |
Weight | One-touch lock type : Approx. 310g Screw soldering type : Approx. 650g |
StandardAccessories
Operation Manual (CD-R), P-57(DC power cable),
P-72(ground wire), Wire connection seals
Bumpon (rubber feet), Screwdriver (for one-touch lock type only)
Terminal block (for screw soldering type only)
Optional Accessories
USB-CAN converter | LEAFLIGHT HS |
Terminating resistance | CANTERM120 |
AC adaptor | SA-10A-EDS |
Connection board//clip | CN-1A |
DIN rail mounting plate |
Y cable | N-103 |
Communication cable 1m | N-102 |
Communication cable 3m | H-11681 |
Communication cable 5m | H-11682 |
Communication cable 10m | H-11683 |
※(NOTE) Please contact our sales department
for communication cables other than those listed above.
Network Data Collector
NTB-201A |
Number of measuring channel | 4 |
Scanning speed | Approx. 0.5 sec/channel |
Applicable sensors | DC voltage-output, Thermocouple |
Voltage-output measurement
Range | Measuring range | Resolution | Accuracy | Signal Source Resistance |
---|
10V | 0~±10.0000V | 100μV | ±(0.1% rdg+0.0003V) | Approx. 1MΩ |
50V | 0~±50.000V | 1mV | ±(0.1% rdg+0.003V) | Approx. 1MΩ |
Thermocouple
Type | Measuring range | Resolution | Accuracy | The internal reference junction accuracy |
---|
K | -200.0 to +1230.0℃ | 0.1℃ | -200.0 to -100℃ or less | ±(0.2%rdg+0.3℃) | ±0.5℃ with input terminal temperature balanced in an ambient temperature range of0 to 50℃) ±1.0℃ (with input terminal temperature balanced in an ambient temperature range of -10 to 0℃ orless) |
-100.0 to +1230.0℃ | ±(0.1%rdg+0.2℃) |
T | -200.0 to +400.0℃ | -200.0 to -100℃ or less | ±(0.2%rdg+0.3℃) |
-100.0 to +400.0℃ | ±(0.1%rdg+0.2℃) |
E | -200.0 to +660.0℃ | -200.0 to -100℃ or less | ±(0.2%rdg+0.3℃) |
-100.0~+660.0℃ | ±(0.1%rdg+0.2℃) |
J | -200.0 to +870.0℃ | -200.0 to -100℃ or less | ±(0.2%rdg+0.3℃) |
-100.0 to +870.0℃ | ±(0.1%rdg+0.2℃) |
R | 0.0~ +1760.0℃ | 0.0 to +100℃ or less | ±(0.2%rdg+0.8℃) |
+100.0 to +1760.0℃ | ±(0.125%rdg+0.6℃) |
N | -200.0~ +1300.0℃ | -200.0 to -100℃ or less | ±(0.2%rdg+0.3℃) |
-100.0 to +1300.0℃ | ±(0.1%rdg+0.2℃) |
(NOTES)1 . Accuracies do not include the internal reference junction compensator accuracy
2. The reference junction compensator is switchable between internal and external
3. The thermocouple resistance should be 1kΩ or less
Check function | Burnout check. |
Power save mode | Provided ON/OFF using “OPT.3” DIP switch. |
Interface | Dedicated interface conforming to CAN, |
Operating temperature range | -10 to 50℃ |
Operating humidity range | 20 to 85%RH (no condensation) |
Power supply | 11 to 16VDC |
Current consumption (during operation at 12VDC) | Constant-voltage bridge excitation Operation : 100mA or less Standby : 60mA or less Standby (in power save mode) : 40mA or less Constant-current bridge excitationOperation : 70mA or less Standby : 60mA or less Standby (in power save mode) : 40mA or less |
Dimensions | One-touch lock type : 150 (W)×28 (H) × 55 (D) mm (excluding protrusions) Screw soldering type : 150 (W)×28 (H) × 110(D) mm (excluding protrusions) |
Weight | Approx. 320g |
(NOTE) TEDS function is unusable
Network Terminal Box Control Software
NTB-10A |
NTB-10A | For remote control of network terminal boxes from a PC, and displaying measurement data in graphs or a numeric format on the PC screen. |
Specifications | Number of measuring units: NTB-100A series: 1 to 99 (The number of MAX CHs is 396.) |
Measuring function | Relative (relative value) measure data with an initial unbalance value deducted. ZERO measurement: Measure initial unbalance value. |
CH conditions | Meas channel ON/OFF, CAL coeffcient calculation ON/OFF, Relative measurement ON/OFF, CAL coeffcient, Offset, Unit, Dec, digits, Ref resist, CH Name (20 characters) |
Meas condition file | Load and save |
Meas operation | MONITOR Meas:Measure ZERO value during MONITOR measurement INTERVAL Meas, File dividing function of the measured data: Not divide,every hour, every day. |
Interval meas | Interval Start Time, Interval Time, Repeat 0 to 9999 (0 to infinite) The number of interval measuring steps: 5 |
Display switching function | List only Numeric display: Arbitrary 1 CH to MAX 5 CHs. Font size (Large, Medium, Small) |
Graph display function of measured data | Graph type : Y-YIME, BAR graph(with perk hold function) With the auto scale function during the measurement |
Measured data saving function | The measured data is saved with the CSV format. |
TEDS compatible | Load the information of the TEDS sensor and automatically set it to the CH condition, CH Name writable to the TEDS information (KYOWA sensor only, within 28 characters) |
Dimension setting of the recorded data | Saved in the hard disk of the PC. |
File split | No split Split every hour Split every day |
System Requirements | CPU:Pentium4 2GHz or higher |
OS:Microsoft Windows XP (32-bit) Microsoft Windows Vista (32-bit) |
Memory :1 GB or more 2 GB or more for Vista |
VistaHard disk:At least 10 MB free disk space (Not including the size of data file to be created) |
Display resolution:1024 × 768 minimum |
Colors:Full color or higher |
USB-CAN converter:Model LEAFLIGHT HS |
Handy Data Logger
SME-100A/101A |
Number of measuring channels | 1 (in independent use of the logger) Max. 33 channels with NTBs connected to the logger*33 channels = 1 + 32 from 8 NTBs |
Sampling frequency | (In independent use, or NTB-dependent when connected to NTBs) Approx. 0.5 sec : 0 to ± 30000 μm/m Approx 1 sec: ± 30000 μm/m or more : Temperature measurement with civil engineering transducer with temperature measuring function |
Measurement mode | RELATIVE mode (the zero value is subtracted from measurements) *“Zero” denotes the initial unbalance during strain measurement, and can be acquired at any time. |
Arithmetic operation | Calculation using a coefficient |
Applicable sensor | Strain gages, strain gage transducers, civil engineering transducers with temperature measuring function Strain Gage | Resistance strain gauge application |
---|
1-gage method | 120,240,350Ω | 2,4gage method | 120 to 1000Ω |
|
120 to 1000Ω | Constant-voltage bridge excitation: Approx. 2VDC Constant-current bridge excitation: Approx. 5.6mA (bridge resistance 350Ω) |
Measuring range | Strain measurement 0 to ± 300000 μm/m (constant-voltage bridge excitation) 0 to ± 20000 μm/m (constant-current bridge excitation) Temperature measurement with civil engineering transducer with temperature measuring function -30.0℃to 70.0℃ |
Resolution | Strain measurement 0 to ± 30000 μm/m : 1 μm/m ± 30000 to ± 300000 μm/m : 10 μm/m Temperature measurement with civil engineering transducer with temperature measuring function 0.1℃ |
Accuracy (NDIS one-touch connector, 4-gage connection) | Strain measurement 0 to ± 30000 μm/m : ± (0.05% rdg. + 2) μm/m ± 30000 to ± 300000 μm/m : ± (0.1% rdg. + 20) μm/m Temperature measurement with civil engineering transducer with temperature measuring function ± 0.5℃ |
Check function | Insulation resistance measurement : 2 to 100MΩ Resistance measurement : 0 to 20kΩ |
Interval measurement | 1 minute to 99 hours 59 minutes in 1-minute steps Starting date/time: year/month/day/hour/minute |
Storage | SD card (optional) |
Applicable SD card | 256MB, 512MB, 1GB, 2GB (FAT16) (SDHC is not applicable) |
Display | Monochrome LCD, 128 × 64 dots |
TEDS | Read from TEDS sensors Operator input of channel names (Kyowa ID only in up to 10 characters) |
Operating temperature & humidity range | -10 to 50℃, 20 to 85%RH (no condensation) |
Power supply | AA battery × 2 Consecutive operation time: Approx. 10 hours (with alkaline batteries, NTB not connected) *Nickel hydride batteries can also be used. *An AC adapter (optional, DR-523E) is provided for SME-101A |
Auto Power Off | Power is automatically turned off if no key operation is detected for 5 minutes. In interval measurement mode with an interval of 3 minutes or longer, power is automatically turned off during standby period and turned on again 1 minute before the next measurement is started (ON/OFF of Auto Power Off can be specified) |
Dimensions | 108.4×188× 41mm (excluding protrusions) |
Weight | Approx. 450g (excluding batteries) |
Standard Accessories
Operation Manual (CD-R), input cable, communication
cable, AA alkali battery × 2, shoulder belt, hand strap